EasyManua.ls Logo

Endress+Hauser Proline Promag W 400 - Page 12

Endress+Hauser Proline Promag W 400
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag W 400
12 Endress+Hauser
EtherNet/IP
Protocol The CIP Networks Library Volume 1: Common Industrial Protocol
The CIP Networks Library Volume 2: EtherNet/IP Adaptation of CIP
Communication type 10Base-T
100Base-TX
Device profile Generic device (product type: 0x2B)
Manufacturer ID 0x49E
Device type ID 0x1067
Baud rates Automatic ¹⁰⁄₁₀₀ Mbit with half-duplex and full-duplex detection
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Supported CIP connections Max. 3 connections
Explicit connections Max. 6 connections
I/O connections Max. 6 connections (scanner)
Configuration options for
measuring device
DIP switches on the electronics module for IP addressing
Manufacturer-specific software (FieldCare)
Add-on Profile Level 3 for Rockwell Automation control systems
Web browser
Electronic Data Sheet (EDS) integrated in the measuring device
Configuration of the EtherNet
interface
Speed: 10 MBit, 100 MBit, auto (factory setting)
Duplex: half-duplex, full-duplex, auto (factory setting)
Configuration of the device
address
DIP switches on the electronics module for IP addressing (last octet)
DHCP
Manufacturer-specific software (FieldCare)
Add-on Profile Level 3 for Rockwell Automation control systems
Web browser
EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
Device Level Ring (DLR) No
Fix Input
RPI 5 ms to 10 s (factory setting: 20 ms)
Exclusive Owner Multicast Instance Size [byte]
Instance configuration: 0x68 398
O → T configuration: 0x66 56
T → O configuration: 0x64 32
Exclusive Owner Multicast Instance Size [byte]
Instance configuration: 0x69 -
O → T configuration: 0x66 56
T → O configuration: 0x64 32
Input only Multicast Instance Size [byte]
Instance configuration: 0x68 398
O → T configuration: 0xC7 -
T → O configuration: 0x64 32
Input only Multicast Instance Size [byte]
Instance configuration: 0x69 -
O → T configuration: 0xC7 -
T → O configuration: 0x65 32

Table of Contents

Other manuals for Endress+Hauser Proline Promag W 400

Related product manuals