Proline Promass E 100
48 Endress+Hauser
Mounting Safety Barrier
Promass 100
> 22.5
(> 0.89)
> 22.5
(> 0.89)
A0016894
23 Minimum distance between additional Safety Barrier Promass 100 or other modules.Engineering unit mm
(in)
Environment
Ambient temperature range
Measuring device • –40 to +60 °C (–40 to +140 °F)
•
Order code for "Test, certificate", option JM:
–50 to +60 °C (–58 to +140
°F)
Safety Barrier Promass 100 –40 to +60 °C (–40 to +140 °F)
‣
If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
Storage temperature
–40 to +80 °C (–40 to +176 °F), preferably at
+20 °C (+68 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Degree of protection
Transmitter and sensor
• As standard: IP66/67, type 4X enclosure
• With the order code for "Sensor options", option CM
: IP69 can also be ordered
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Safety Barrier Promass 100
IP20
Vibration resistance
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz
, 3.5 mm peak
– 8.4 to 2 000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g
2
/Hz
– 200 to 2 000 Hz, 0.001 g
2
/Hz
– Total: 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Impact resistance
Rough handling shocks according to IEC 60068-2-31