Proline Promass E 100
10 Endress+Hauser
Assignable measured
variables
Mass flow
Volume flow
Corrected volume flow
Density
Reference density
Temperature
The range of options increases if the measuring device has one or more
application packages.
Switch output
Switching behavior Binary, conductive or non-conductive
Switching delay Adjustable: 0 to 100 s
Number of switching
cycles
Unlimited
Assignable functions Off
On
Diagnostic behavior
Limit value
Mass flow
Volume flow
Corrected volume flow
Density
Reference density
Temperature
Totalizer 1-3
Flow direction monitoring
Status
Partially filled pipe detection
Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
PROFIBUS DP
Signal encoding NRZ code
Data transfer 9.6 kBaud…12 MBaud
Modbus RS485
Physical interface In accordance with EIA/TIA-485-A standard
Terminating resistor For device version used in non-hazardous areas or Zone 2/Div. 2: integrated
and can be activated via DIP switches on the transmitter electronics module
For device version used in intrinsically safe areas: integrated and can be
activated via DIP switches on the Safety Barrier Promass 100
EtherNet/IP
Standards In accordance with IEEE 802.3
PROFINET
Standards In accordance with IEEE 802.3
Signal on alarm
Depending on the interface, failure information is displayed as follows: