EasyManua.ls Logo

Endress+Hauser Proline Promass E 100 - Ex Connection Data

Endress+Hauser Proline Promass E 100
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass E 100
12 Endress+Hauser
Local display
Plain text display With information on cause and remedial measures
Backlight Red backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Interface/protocol
Via digital communication:
HART protocol
PROFIBUS DP
Modbus RS485
EtherNet/IP
PROFINET
Via service interface
CDI-RJ45 service interface
Plain text display With information on cause and remedial measures
Additional information on remote operation →  79
Web server
Plain text display With information on cause and remedial measures
Light emitting diodes (LED)
Status information Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
Supply voltage active
Data transmission active
Device alarm/error has occurred
EtherNet/IP network available
EtherNet/IP connection established
PROFINET network available
PROFINET connection established
PROFINET blinking feature
Ex connection data
These values only apply for the following device version:
Order code for "Output", option M "Modbus RS485", for use in intrinsically safe areas
Safety Barrier Promass 100
Safety-related values
Terminal numbers
Supply voltage Signal transmission
2 (L-) 1 (L+) 26 (A) 27 (B)
U
nom
= DC 24 V
U
max
= AC 260 V
U
nom
= DC 5 V
U
max
= AC 260 V