Proline Promass E 100
Endress+Hauser 9
Maximum possible full scale value:
max(G)
=
max(F)
· ρ
G
: x = 70 000
kg/h · 60.3 kg/m³ : 125 kg/m³ = 33 800 kg/h
Recommended measuring range
"Flow limit" section → 53
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value do not override the electronics unit, with the result that
the totalizer values are registered correctly.
Output
Output signal HART current output
Current output 4-20 mA HART (active)
Maximum output values • DC
24 V (no flow)
• 22.5 mA
Load 0 to 700 Ω
Resolution 0.38 µA
Damping Adjustable:
0.07 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Function Can be set to pulse, frequency or switch output
Version Passive, open collector
Maximum input values • DC 30 V
• 25 mA
Voltage drop For
25 mA: ≤ DC 2 V
Pulse output
Pulse width Adjustable: 0.05 to 2 000 ms
Maximum pulse rate 10 000 Impulse/s
Pulse value Adjustable
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
Frequency output
Output frequency Adjustable: 0 to 10 000 Hz
Damping Adjustable: 0 to 999 s
Pulse/pause ratio 1:1