Proline Promass E 100
8 Endress+Hauser
Input
Measured variable Direct measured variables
Mass flow
Density
Temperature
Calculated measured variables
Volume flow
Corrected volume flow
Reference density
Measuring range Measuring ranges for liquids
DN Measuring range full scale values 
min(F)
to 
max(F)
[mm] [in] [kg/h] [lb/min]
8 ³⁄₈ 0 to 2 000 0 to 73.50
15 ½ 0 to
6 500 0 to 238.9
25 1 0 to 18 000 0 to 661.5
40 0 to 45 000 0 to 1 654
50 2 0 to 70 000 0 to 2 573
80 3 0 to 180 000 0 to 6 615
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the formula below:
max(G)
= 
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
< 
max(F)
max(G)
can never be greater than 
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x Constant dependent on nominal diameter
DN x
[mm] [in] [kg/m
3
]
8 ³⁄₈ 85
15 ½ 110
25 1 125
40 125
50 2 125
80 3 155
To calculate the measuring range, use the Applicator sizing tool →  88
Calculation example for gas
Sensor: Promass E, DN 50
Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
Measuring range (liquid): 70 000 kg/h
x = 125 kg/m³ (for Promass E, DN 50)