Device Functions Proline t-mass 65 16 Group PROCESS PARAMETER
Endress+Hauser 57
REFERENCE TEMPERA-
TURE
Use this function to enter a value for reference temperature with which the reference
density is calculated for the corrected volume flow measurement.
User input:
5-digit floating-point number
Factory setting:
0.0000 [°C] or +32.000 [°F]
(country dependent, → Page 71 ff.)
! Note!
The appropriate unit is taken from the function UNIT TEMPERATURE in the function
group SYSTEM UNITS (see Page 11).
REFERENCE PRESSURE Use this function to enter a value for reference pressure with which the reference density
is calculated for the corrected volume flow measurement.
User input:
5-digit floating-point number
Factory setting:
1.0130 [bar a] or 14.692 [psi a]
(country dependent, → Page 71 ff.)
! Note!
The appropriate unit is taken from the function UNIT PRESSURE in the function group
SYSTEM UNITS (see Page 11).
REFERENCE DENSITY This function displays the calculated reference density used for the corrected volume flow
measurement.
Display:
5-digit floating-point number
Factory setting:
1.2930 [kg/m
3
] or 0.0807 [lb/ft
3
]
(country dependent, → Page 71 ff.)
! Note!
The appropriate unit is taken from the function UNIT DENSITY in the function group
SYSTEM UNITS (see Page 11).
Function description PROCESS PARAMETER