9 Trouble-shooting Promass 40
36 Endress+Hauser
9.2 System error messages
Caution:
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. The procedures on Page 6 must be carried out before you return a flowmeter
to Endress+Hauser.
Always enclose a duly completed “Safety regulation” form. You will find a preprinted
form at the back of this manual.
Type Error message / No. Cause Remedy / spare part
Serious system errors are always recognized by the instrument as “Fault message”, and are shown as a
lightning flash on the display. Fault messages immediately affect the inputs and outputs.
Simulation and positive zero return, on the other hand, are classed and displayed as notice messages.
Also observe the informations on
→ Page 26 ff. and 42.
S = System error
= Fault message (with an effect on the inputs and outputs)
! = Information message (without an effect on the inputs and outputs)
S
CRITICAL FAILURE
# 001
Serious device error Replace the amplifier board. Spare
parts
→ Page 44
S
AMP HW EEPROM
# 011
Amplifier:
Defective EEPROM
Replace the amplifier board. Spare
parts
→ Page 44
S
AMP SW EEPROM
# 012
Amplifier:
Error accessing EEPROM data
The EEPROM data blocks in which
an error has occurred are dis-
played in the “RESTORE DATA
FAILURE” function.
Press Enter to acknowledge the
errors in question; default values
are automatically inserted instead
of the errored parameter values.
Note:
The measuring device has to be
restarted if an error has occurred
in a totalizer block (see error No.
111 / CHECKSUM TOTAL).
S
SENSOR HW DAT
# 031
Sensor:
1. S-DAT™ is defective.
2. S-DAT™ is not plugged into
the I/O board or is missing.
1. Replace the S-DAT™.
Spare parts
→ Page 44
Check the spare part set
number to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
2. Plug the S-DAT™ into the
I/O board
→ Page 46