Promass 40 10 Technical data
Endress+Hauser 55
10.1.9 Operating conditions (process)
Medium temperature
range
Sensor:
40...+125 °C
Seals:
no internal seals
Fluid pressure limit
(nominal pressure)
Flanges: DIN PN 40...100 / ANSI Cl 150, Cl 300, Cl 600 / JIS 10K, 20K, 40K, 63K
The Promass E sensor does not have a secondary containment
Limiting flow See Page 50 ff. (Measuring range)
Select nominal diameter by optimising between required flow range and permissible
pressure loss. See Page 50 ff. for a list of maximum possible full scale values.
The minimum recommended full scale value is approx.
1
/
20
of the maximum full scale
value.
In most applications, 20...50% of the maximum full scale value can be considered
ideal.
Select a lower full scale value for abrasive substances such as fluids with entrained
solids (flow velocity < 1 m/s).
For gas measurement the following rules apply:
Flow velocity in the measuring tubes should not be more than half the sonic velocity
(0.5 Mach).
The maximum mass flow depends on the density of the gas (see formula on
Page 50).