Promass 40 10 Technical data
Endress+Hauser 49
10 Technical data
10.1 Technical data at a glance
10.1.1 Applications
The measuring device is for mass flow and volume flow measurement of liquids and
gases in sealed piping systems. Application examples:
additives
oils, fats
acids, alkalis
lacquers, paints
suspensions
gases.
10.1.2 Function and system design
Measuring principle Mass flow measurement by the Coriolis principle
Measuring system The measuring system consists of a transmitter and a sensor:
Promass 40 transmitter
Promass E sensor
10.1.3 Input
Measured variable Mass flow (proportional to the phase difference between two sensors mounted on the
measuring tube to register a phase shift in the oscillation)
Volume flow (calculated from mass flow and fluid density, which is proportional to the
resonance frequency of the measuring tubes)
Fluid temperature (registered by temperature sensors) for compensating temperature
effects.