Installation Solicap M FTI55, FTI56
20 Endress+Hauser
3.8 Rope probes FTI56
!
Note!
Total length of the probe from the start of the thread: L = L1 + L3
Rope probe
Partially insulated rope
Rope probe
Fully insulated rope
Rope probe
with inactive length
Partially insulated rope
Rope probe
with inactive length
Fully insulated rope
TI418F38
TI418Fen40
H2 66 66 66 66
H3 25 25 25 25
Across flats (AF) 50 50 50 50
Total length (L) 500...20000 500...20000 700...20000 700...20000
Active rope length (L1) 500...20000 500...20000 500...19800 500...19800
Length of partial insulation (L2)* 500 - 500 -
Inactive length (L3) - - 200...2000 200...2000
ø Inactive length - - 43 43
Probe rope diameter
(with insulation)
6
(8)
12
(14)
6
(8)
12
(14)
6
(8)
12
(14)
6
(8)
12
(14)
ø Tensioning weight** 30 40 30 40 30 40 30 40
Length of tensioning weight (lg) 150 250 150 250 150 250 150 250
Tensile loading capacity (kN) of probe
rope at 20 °C
30 60 30 60 30 60 30 60
Maximum process temperature (°C) 180 120 180 120
For use in mounting nozzles - - X X
In the event of condensate on tank
ceiling
--XX
X = recommended
* The length of the partial insulation extends, at maximum, to the tensioning weight.
** The tensioning weight is always uninsulated.