Solicap M FTI55, FTI56 Installation
Endress+Hauser 29
3.10 Installation
3.10.1 Probe with thread
R 1½ and 1½ NPT (conical):
Where necessary, wrap sealing material around the thread. Ensure that the electrical connection
between the probe and the tank is correct.
If the process connection of the probe is insulated from the metal tank (e.g. using seal material),
the ground connection on the probe housing must be connected to the tank using a short line.
"
Caution!
Do not damage the probe insulation during installation.
Do not turn the housing while screwing in the probe, as otherwise the housing fixture can be
damaged.
BA300Fxx007
3.10.2 Installation tools
The following tools are required for installation:
Tool for mounting flanges
or a size 50 Allen key for the threaded connection
and a Phillips-head screwdriver for aligning the cable entry.