Solicap M FTI55, FTI56 Installation
Endress+Hauser 35
3.14.2 Wall mounting
Push the bracket onto the sleeve and screw it into place.
Mark the distance between the holes on the wall, and then drill the holes.
Screw the separate housing to the wall.
BA300Fxx010
3.14.3 Pipe mounting
Push the bracket onto the sleeve and screw it into place.
Screw the separate housing to the pipe (max. 2").
BA300Fxx011
3.15 Post-installation check
After installing the measuring device, carry out the following checks:
Is the device damaged (visual inspection)?
Does the device correspond to the measuring point specifications, including process temperature
and pressure, ambient temperature, measuring range, etc.?
Is the process connection tightened with the correct torque?
Are the measuring point number and labeling correct (visual inspection)?
Is the measuring device adequately protected from precipitation and direct sunlight?