Solicap M FTI55, FTI56 Operation
Endress+Hauser 51
5.2 Human interface and display elements for
FEI53, FEI57S
The electronic inserts FEI53 and FEI57S are used in conjunction with Nivotester switching devices.
The functions of the DIP switches (A and B) and the LEDs (1 and 2) are described in the table below.
The operating status of the device is indicated by LEDs (LED 1 and 2) on the electronic insert and
provides information on operational readiness (1) and, where applicable, the type of fault (2).
BA300Fen016
LED 1 operational : Flashes at 5-second intervals.
LED 2 fault : The red LED flashes if there is a fault that you can correct.
LED 2 fault : The red LED lights up continuously if the device has a fault that cannot be corrected. See also
Page 78, "Troubleshooting".
!
Note!
A description of the human interface and display elements of the Nivotester switching device is
provided in the documentation that accompanies the device.
DIP switch Function
A
Standard
1)
: If the measuring range is exceeded no alarm is output.
A
: If the measuring range is exceeded an alarm is output.
B
Measuring range: The mesasuring range is between 0 to 500 pF.
Span: The span is between 5 to 500 pF.
B
Measuring range: The mesasuring range is between 0 to 1600 pF.
Span: The span is between 5 to 1600 pF.