Installation Solicap M FTI55, FTI56
32 Endress+Hauser
3.12.1 Shortening the connecting cable
A recalibration must be performed before commissioning ä 53
!
Note!
The maximum connection length between the probe and the separate housing is 6 m. When
ordering a device with a separate housing, the desired length must be specified.
If the connecting cable is to be shortened or guided through a wall, it must be disconnected at the
process connection. To do so, proceed as follows:
Unscrew the pressing screw (1) using a 22mm open-end wrench. If necessary, hold the process
connection. In doing so, ensure that neither the connecting cable nor the probe is turned in the
process.
Pull the insert seal (2) out of the cable gland (3).
Using a 22mm open-end wrench, disconnect the cable gland (3) from the adapter disk. If
necessary, hold it against the adapter disk (6) using a 34mm open-end wrench.
Disconnect the adapter disk (6) from the adapter bushing (18).
Remove the snap ring with a snap ring pliers.
Grip the nut (M6) on the blade plug with a pliers and pull out the blade plug.
Then, shorten the connecting cable to the desired length.
If the separate housing has to be mounted in a different room than the probe, you can now route
the connecting cable through the wall.
You can now reassemble the device by following the reverse order of steps.
!
Note!
If you shorten the connecting cable, we recommend reusing all strands with ring terminals.
If the strands are not to be reused, the crimp connections of the new ring terminals attached must
be insulated with a heat-shrinking sleeve tube, for example (danger of short circuit).
All soldered joints must be insulated. Use heat-shrinking sleeves to do so.