Commissioning Solicap M FTI55, FTI56
56 Endress+Hauser
6.2.4 Carrying out full calibration
!
Note!
The full calibration measures the capacitance value of the probe when the tank is full. If the
measured capacitance value is, for example, 100 pF (full calibration), a switching threshold of
2 pF is subtracted from this value. The capacitance value of the switch point is thus 98 pF.
The switching threshold depends on the value set for the switch point adjustment (for more
information, see ä 59).
To carry out a full calibration, proceed as follows:
1. Make sure that the probe is covered by the medium up to the desired switch point.
2. Turn the function switch to position 2.
3. Press the "+" key for at least two seconds.
4. Release the "+" key when the green LED 5 starts to flash.
The process of saving the full calibration is complete when the green LED 5 lights up continuously.
You can turn the function switch back to position 1 to return to operation.
Function switch
position
Function – key + key Light emitting diodes (LED signals)
A B C 1 (green) 2 (green) 3 (red) 4 (green) 5 (green) 6 (yellow)
2
Full calibration Press On
(present)
**
** Switch status signaling (on/off/flashing) depends on the mounting location selected and the failsafe mode (MIN/MAX) set.
The LED flashes if a calibration has not yet been carried out.
1
7
3
5