Commissioning Solicap M FTI55, FTI56
58 Endress+Hauser
6.2.6 Reset: Calibration and switch point adjustment
To reset the calibration/switch-point shift (all the other settings remain unchanged), proceed as
follows:
1. Turn the function switch to position 2.
2. Press both the "–" and "+" keys for at least 10 seconds.
3. The green LEDs 1-5 light up in succession.
The reset calibration has been carried out and saved. The yellow LED 5 flashes.
The device is not operational until you have carried out a new calibration.
The switch point adjustment is reset to the factory setting of 2 pF.
Function switch
position
Function – key + key Light emitting diodes (LED signals)
A B C 1 (green) 2 (green) 3 (red) 4 (green) 5 (green) 6 (yellow)
2
Reset:
Calibration and switch
point adjustment
Press both keys for
approx. 10 s
On –> –> –> –> **
** Switch status signaling (on/off/flashing) depends on the mounting location selected and the failsafe mode (MIN/MAX) set.
The LED flashes if a calibration has not yet been carried out.