Commissioning Solicap M FTI55, FTI56
66 Endress+Hauser
6.2.13 Restoring factory settings
!
Note!
This function allows you to restore the factory settings. This is particularly useful if the device has
already been calibrated once and, for example, there is a fundamental change in the medium in
the tank.
After restoring the factory settings, you must repeat the calibration.
To restore the factory settings, proceed as follows:
1. Turn the function switch to position 1.
2. Press the "+" and "–" keys simultaneously for approx. 20 seconds. During the time it takes to
restore the factory settings, the LEDs 1–5 light up consecutively.
3. The factory settings have been successfully restored if the green LED 1 and the yellow LED are
flashing.
You have now restored the factory settings and can continue with setting the measuring range and
the calibration.
Function switch
position
Function – key + key Light emitting diodes (LED signals)
A B C 1 (green) 2 (green) 3 (red) 4 (green) 5 (green) 6 (yellow)
1
Operation Flashes
Operational
LED
On***
(MIN-SIL)
Flashes
(warning/
alarm)
On***
(MAX-SIL)
On/off/
flashes**
Restore factory setting Press both keys for
approx. 20 s
On –> –> –> –> **
* These settings are factory settings.
** Switch status signaling (on/off/flashing) depends on the mounting location selected and the failsafe mode (MIN/MAX) set.
The LED flashes if a calibration has not yet been carried out.
*** Only in conjunction with electronic insert FEI55 (SIL). The device is in the SIL mode. To change the current settings, the device must be unlocked.
1
7
3
5
_
+