Commissioning Solicap M FTI55, FTI56
70 Endress+Hauser
6.3.2 Setting the measuring range
!
Note!
• The choice of measuring range (0 to 500 pF and 0 to 1600 pF) depends on the function of the
probe.If the probe is used as a limit switch, you can retain the factory setting of 0 to 500 pF.
• If the probe is used for two-point control, the following recommendations apply for vertical
installation:
– Measuring range from 0 to 500 pF for probe lengths up to 1.0 m
– Measuring range from 0 to 1600 pF for probe lengths up to 4.0 m
All other settings must be made on the respective Nivotester switching device.
6.3.3 Output signals
Output signal FEI53
TI418Fen46
* See ä 78, "Troubleshooting"
DIP switch Function
B
Measuring range: The mesasuring range is between 0 to 500 pF.
Span: The span is between 5 to 500 pF.
B
Measuring range: The mesasuring range is between 0 to 1600 pF.
Span: The span is between 5 to 1600 pF.
ModeMode
InstrumentInstrument failurefailure
MaintenanceMaintenance
required *required *
Output signalOutput signal
Normal operationNormal operation
< 2,7 V< 2,7 V
33at terminalat terminal
3...12 V3...12 V
at terminal 3at terminal 3
3...12 V3...12 V
33at terminalat terminal
LEDsLEDs
green redgreen red