11.2.9.a Vacuum pulsed mode
The underlining reason for vacuum pulsation is creating a massage effect during a treatment
session. This is achieved by rhythmically varying the amount of suction under the electrode
cups. To allow pulsation the range must be set between 10-30%. Ideally the setting should be
between 50-70%. The minimum vacuum setting that allows pulsation is 15. If during treatment
the vacuum setting is reduced below 15 then the pulsation range will also automatically be
reduced to insure that the amount of suction does not drop below the minimum needed for
preventing the cup to fall off. The pressure under the cup will rhythmically pulse between the
vacuum setting and the selected range. The rhythm cycle can be adjusted with the parameter
“Hold time” and can be adjusted from 0 to 2.5 seconds in steps of 0.5 seconds.
The hold time setting is applied when the lower vacuum limit is reached. (Please see par.12.13)
11.2.9.b Vacuum electrode selection
• Using the central
controller [5], select
electrode selection.
• Enter the selection with
the confirmation button
[6]
• Select in the pop-up
with the central
controller [5] electrodes
or Vacuum cups.
• Push the confirmation
button to return.