11.2.10.f Template
Select Template with
the central controller [5]
and enter with the
confirmation button [6]
Using the central
controller [5], to select
Threshold and push
the confirmation button
[6].
Adjust:
Template
Template Time in sec
Rest Time in sec
Cycles (amount of
repetitions)
Sensitivity
Template Height
(feedback on screen)
Go to record to start the
treatment.
The template form is
displayed in orange as the
guide line for the patient to
follow.
Note!
Stop the treatment via
the assist button
Use the pause under
the assist button to
interrupt the treatment.
Use the same button to
start again.
Feedback Therapy
4
Myofeedback
Continuous
Exercise / Rest
Template
Continuous 1+2
Exercise / Rest 1+2
Template 1+2
0:00
0:00
E1 10:00
0.0 mA
0.0 mA
Myofeedback
4
Template - Settings
Template
Template Time
0:05
Rest Time
0:05
Cycles
5
Sensitivity
200 µV
Shift
0 µV
Template Height
70 %
0:00
0:00
E1 01:15
0.0 mA
0.0 mA
Myofeedback
4
Template
0:00
0:00
E1 01:05
0.0 mA
0.0 mA
5 µV

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Enraf Nonius 6 Series and is the answer not in the manual?

Enraf Nonius 6 Series Specifications

General IconGeneral
BrandEnraf Nonius
Model6 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals