59
1600751-4D
Treatment and
registration runs
Note!
Following parameters are
adjustable during the
treatment:
Sensitivity
Shift
Threshold
Stop the training via
the assist button or
wait until the training is
finished.
Go one step back and
select analyze or
report.
Store these data via the assist button on a USB stick. This only possible in combination with
the MyoComBox
The final result screen
depends on the
selected exercises
parameters
The parameters
included can be stored
on the internal
memory on the unit.
Graph and figures can
only be exported to an
USB stick via the
Myocombox.
11.2.10.a Connecting to the MyoComBox
A device with a feedback therapy module has some special settings.
In order to be able to work with the MyoComBox the device has to be connected to the
MyoComBox by means of the “System Settings.”
1. Turn on the MyoComBox (make sure that the Bluetooth antenna is in one of the USB
connections of the MyoComBox).
2. Go to the “System Settings” via the Home menu. Select the MyoComBox and then
select “Search.” The device will show the serial numbers of the available
MyoComBox(es) in a list. Connect the device to the MyoComBox of your choice by
selecting the serial number from the list. Check that the serial number corresponds to
the number on the left in the LCD screen of the MyoComBox.
3. Set the time and date of the MyoComBox.
4. Designate a system name if desired. This name will be displayed in the selected
MyoComBox.
In addition to these settings there is also the option of setting a notch filter. The notch filter
removes artifacts from the EMG signal that are the result of the power supply frequency. Set the
Myofeedback
4
Continuous Record
0:00
0:00
E1 09:45
0.0 mA
0.0 mA
85 µV
Myofeedback
Continuous - Report
Minumum
304 µV
Maximum
355 µV
Average
319 µV
Total recording time
0:25
15:00
15:00
0.0 mA
0.0 mA

Table of Contents