11.2.10.g Shift
Shift is a special tool only available in the Myomed 6- series. The function of this parameter is to
zoom-in on a specific area in the graphic to have a better feedback for the patient. This is useful
for patient with muscle Hypertonia.
The adjustment is as follow:
• Define the ideal bandwidth in the sensitivity setting screen.
• Set the sensitivity back to the ideal bandwidth, the signal will disappear in the graphic.
Go back to the setting menu and select Shift.
• Scroll with the shift until the signal is back on the screen. The sensitivity range is much
lower and only for a limited range.
Note!
• The shift range is
linked to the Sensitivity
setting.
11.2.10.h Myofeedback + Stimulation
1. Place the reference electrode on the skin and connect it with the unit.
2. Place the EMG electrodes on the skin and connect these with the unit.
!CAUTION
• Connection of accessories other than the ones specified by the manufacturer can
adversely affect the safety of the patient and correct functioning of the equipment, and is
therefore not permitted.