83
1600751-4D
The StatUS applicator / ultrasound treatment head may be disinfected with a moist cloth dipped
in 70% alcohol solution.
Cable/Rubber Cup
We recommend cleaning the cable and the rubber cup daily with a moist cloth. Use lukewarm
water and a little household cleanser if necessary (no abrasives or no alcohol solutions).
Note!
Do not rinse the StatUS applicator / ultrasound treatment head under running water!
Never immerse the StatUS applicator / ultrasound treatment head in water or other
(cleaning) liquids!
For proper cleaning of the StatUS applicator / ultrasound treatment head’s contact
surface, the rubber cuff may be folded back on one movement (see Figure A).
If necessary, you can push up the StatUS applicator / ultrasound treatment head a bit from
beneath (see Figure B).
ATTENTION: Never fold the entire cuff ! (See Figure C.)
Note:
In order to guarantee good transmission of the ultrasound energy, you must use original Enraf-
Nonius gelpads.
Do not use regular (liquid) sonogel because this will be sucked up into the vacuum pump!
Warranty claims in which the remnants of regular ultrasound gel are found in the pump, hoses
or connections will be rejected by Enraf-Nonius.
Figure A