78
Immunity Test
IEC 60601 level
Level of
compliance
Guidance for EM environment
Voltage Interruption
IEC 61000-4-11
0 % UT
for 250 cycles at
50 Hz
for 300 cycles at
60 Hz
0 % UT
for 250 cycles at
50 Hz
for 300 cycles at
60 Hz
Power
frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
for home health care environment
and a professional health care
establishment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
6 V in ISM band
and
from 0.15 MHZ
to 80 MHZ,
amateur radio
band included
80% MA at 1
KHz
3 Vrms
150 kHz to 80
MHz
6 V in ISM band
and
from 0.15 MHZ
to 80 MHZ,
amateur radio
band included
80% MA at 1
KHz
for home health care environment
and a professional health care
establishment
Electromagnetic fields
Radiated RF*
IEC 61000-4-3
10 V/m
80 MHz to 2.5
GHz
10 V/m
80 MHz to 2.5
GHz
for home health care environment
and a professional health care
establishment

Table of Contents