Immunity Test
IEC 60601 level
Level of
compliance
Guidance for EM environment
Proximity fields emitted
by RF wireless
communication devices
IEC 61000-4-3
(provisional method)
9 V/m
:
710
MHz, 745 MHz,
780 MHZ, 5240
MHz, 5550 MHz,
5785 MHz
27 V/m : 385
MHz
28 V/m: 450
MHz, 810 MHz,
870 MHz, 930
MHz, 1720 MHz,
1845 MHz, 1970
MHz, 2450 MHz
9 V/m
:
710
MHz, 745 MHz,
780 MHZ, 5240
MHz, 5550 MHz,
5785 MHz
27 V/m : 385
MHz
28 V/m: 450
MHz, 810 MHz,
870 MHz, 930
MHz, 1720 MHz,
1845 MHz, 1970
MHz, 2450 MHz
for home health care environment
and a professional health care
establishment
Recommended separation distance
EO150 device must be used in an electromagnetic environment in which the perturbations due to RF
are controlled.
The user or installer of the device can help to prevent any electromagnetic interference by maintaining
a minimum distance depending on RF emitter maximum power. Portable RF devices (included cables
and antennas) must not be used closer than 30 cm (12 inches) from any part of the EO150, including
specified cables. Not respecting this recommendation could lead to performance alteration.
NOTE:
Additional technical description (pneumatical description, theory of operation,
measurement uncertainty, functional tests) can be found in the technical manual