41
Instructions for hygienic reprocessing at patient change
The following process must be followed before change of patient:
Wipe disinfection (of device housing exterior)
Replacement of Bacterial filter or HME filter
Replacement of patient circuit or sterilization of reusable circuit system
Replace Air filter
Function Check
Follow this procedure also for devices which have been previously been used by patients in whom
MRSA infection, for example, has been verified. Please take all precautions for your own protection
when reprocessing a device.
For disinfection, we recommend usage of products such as Mikrozid® sensitive liquid from Schülke
or WILAsil® from WILAmed. The following basic products could be used:
- IPA (Iso Propyl Alcohol)
- Ethanol 70%
- Hydrogen peroxyde
- Chlore (1000 ppm)
For other product usage, please contact our customer service.
It is important to avoid the spilling of liquid on the electrical connector of the device. User must
ensure that the disinfectant is dry before turning on the device. It is possible to use other
equivalent disinfectants, however, the proof of compatibility of the product with the device is
the responsibility of the manufacturer of this product.
The EO-70 device external surfaces in contact with the disinfectant products are made with the
materials listed below:
- PC ASA UL94V0
- Terluran GP22 (ABS)
- FT Santoprene 8211-55B100
- Glass
- Polyester
- Polypropylene
SMD Module
Air filter
Air filter cover