39
Internal battery run time is determined by:
Environmental conditions (operating conditions See Technical Specifications)
The condition and age of the battery
The device settings
The patient circuit used and the involuntary leaks
Storing and recharging
If the device is being stored with a 100% charged battery, its internal battery must be recharged
every 4 months maximum during the storage period (recommended 2 months).
Never store a device with a depleted battery.
Prepare the battery for long-term storage
1. The battery charge level should be 100%.
2. Turn off the device.
3. Remove the power cord from the device.
Travelling with EO-70
The EO-70 must be transported in its transport bag.
device should not be operated while in the Transport.
Do not place any heavy or bulky objects in the
zippered pocket on the inside front of the
bag. This could result in damage to the touch screen.
Chapter 4 – Alarms indicator
The EO-70 is equipped with an alarm indicator to alert the user if the device is not functional for a
technical reason. Those alarms are low priority (yellow indicator with no sound) and will all
necessitate a technical support.
If the yellow alarm indicators are active, call your service provider or Eove customer support to have
your device serviced.