40
Chapter 5 - Routine Cleaning and Maintenance
Patients are vulnerable to infections. All equipment shoul
disinfected.
Keep the device, and accessories away from water and all other liquids not specified in this
document. Always turn off and unplug the device before cleaning and verify that it is dry
before plugging it back in.
Clean only exterior surfaces of the
If necessary, wipe the exterior of the device with a damp cloth using a mild cleaning solution.
components and hoses, follow the manufacturer's recommendations for
cleaning and maintenance.
For disinfection, we recommend usage of products such as Microzid
Schülke or WILAsil® from WILAmed. For other product usage, please contact our customer
service.
Warning
Proper cleaning and maintenance of your EOVE device is essential. Cleaning described in this section
should be carried out regularly.
Refer to the user guides of any accessories in use for detailed instructions specific to those devices.
Maintenance Method Frequency
connections and circuit
adapters for any moisture or
contaminants
necessary using appropriate
cleaning solutions
internal battery
external power and operate it
for 10 minutes minimum.
2. Check the remaining battery
level.
3. Restore the mains when the
test is done.
Every six months (with normal
daily use)
air filters cannot be washed or reused.