Chapter 5 - Routine Cleaning and Maintenance
WARNING
Patients are vulnerable to infections. All equipment shoul
d be
regularly cleaned and
disinfected.
Keep the device, and accessories away from water and all other liquids not specified in this
document. Always turn off and unplug the device before cleaning and verify that it is dry
before plugging it back in.
CAUTION
Clean only exterior surfaces of the
EO
-
70
device.
If necessary, wipe the exterior of the device with a damp cloth using a mild cleaning solution.
For all circuit
components and hoses, follow the manufacturer's recommendations for
cleaning and maintenance.
For disinfection, we recommend usage of products such as Microzid
®
sensitive liquid from
Schülke or WILAsil® from WILAmed. For other product usage, please contact our customer
service.
Warning
Proper cleaning and maintenance of your EOVE device is essential. Cleaning described in this section
should be carried out regularly.
Refer to the user guides of any accessories in use for detailed instructions specific to those devices.
Maintenance Method Frequency
Inspect the condition of
the
connections and circuit
adapters for any moisture or
contaminants
Replace and clean as
necessary using appropriate
cleaning solutions
Weekly
Check the charge
level of the
internal battery
1.Unplug the device from
external power and operate it
for 10 minutes minimum.
2. Check the remaining battery
level.
3. Restore the mains when the
test is done.
Every six months (with normal
daily use)
CAUTION
The
air filters cannot be washed or reused.