Guidance and Manufacturer’s Declaration Electromagnetic Emissions & Immunity
WARNING
The
EO
-
70
should not be used in close proximity to
other equipment or stacked on top of
other devices. If this kind of use is unavoidable, the device should be checked carefully and
observed to ensure correct functioning of the device.
Only accessories recommended for the
EO
-
70
should be used
. Using any other accessories
could result in risk to the device or the patient.
Any additional equipment connected to medical electrical de
vices must comply with the
respective IEC or ISO standards (eg, IEC 60950 for data processing equipment).
Furthermore all configurations shall comply with the requirements for medical
electrical systems (see IEC 60601-1-1 or clause 16 of the 3Ed. of IEC 60601-1, respectively).
Adding additional equipment configures a medical system and this system must comply
with the requirements for medical electrical systems. Any person undertaking this kind of
addition shall be responsible to ensure that all requirements are complied with. It is
important to note that local laws take priority over the above mentioned requirements. If
in doubt, consult an EOVE representative or the technical service department.
Interference may occur in the vicinity of equipment marked with the following symbol:
EO
-
70
is designed for use in the electromagnetic environment described below. Those using
the device should ensure that the EO-70 is used in such an environment.
No RF communication equipment (including ante
nna’s cables or antennas) should be closer
than 30 cm (12 inch) of any part of EO-70, including cables. Otherwise the functioning of
these devices could be altered.
Electromagnetic emissions
Emissions test
Level of compliance
Guidance for EM e
nvironment
RF emissions CISPR 11
Class B
EO
-
70
is suitable for home health
care environment and a professional
health care establishment
Harmonic Emissions
IEC 61000-3-2
C
lass A
Voltage Fluctuations/Flicker
Emissions
IEC 61000-3-3
Complying