Electromagnetic immunity
Immunity Test
IEC 60601 level
Level of compliance
Guidance for EM environment
Electrostatic discharge
(ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
F
or home hea
lth care
environment and a
professional health care
establishment
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output
lines
2 kV for power
supply lines
2 kV for
input/output lines
F
or home health care
environment and a
professional health care
establishment
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
mode
1 kV differential
mode
2 kV common mode
F
or home health care
environment and a
professional health care
establishment
Voltage dips, short
interruptions and
voltage
variations on power
supply input lines
IEC 61000-4-11
0% Ut for 0.5
cycle
With 0°, 45°,
90°, 135°, 180°,
225°, 270° et
315°
0% Ut for 1
cycles
70% Ut for 25
cycles at 50 Hz
0% Ut For 30
cycles at 60 Hz
Monophased at
0% Ut for 0.5
cycle
With 0°, 45°, 90°,
135°, 180°, 225°,
270° et 315°
0% Ut for 1 cycles
70% Ut for 25 cycles
at 50 Hz
0% Ut For 250 cycles
at 60 Hz
Monophased at 0°
Mains power quality should be
as home health care
environment and a
professional health care
establishment
If operating during power
cuts, it is recommended to use
other power sources.