43
Essential performances
The EO-70 essential performances can be evaluated with this reference test :
Settings:
INEX Mode, Operating mode: AUTO, Inh. Pressure: 30 cmH2O, Slope: 3, Inh. Time: 3 s, Trigger : off,
Pause : 1 s, Peep: 4 cmH2O, Exh. Pressure: - 30 cmH2O, Exh Time : 2 s, Insp. Oscillations I: on (Amp.
2, Freq. 10 Hz), Exp. Oscillations : OFF.
In these conditions, with a 22 mm tubing and a test lung (Compliance : 20 mL/mbar, Resistance : 20
mbar / L / s, Max volume 1000 mL), monitoring should be the following:
Tidal Volume: 700 ml +/- 140 ml
Peak Flow: 125 l/min +/- 25 l/min