5
effects of hyper- and hypoglycemia and
how to prevent these conditions.
If the pump fails to deliver your needed
insulin, immediately stop using the system
and consider using your emergency kit
to supplement insulin. Contact your
Please be sure to operate in strict
accordance with this user manual,
as failure to follow the instructions
This product can only be used to
administer U-100 insulin.
Only MicroTech Medical disposable
supplies should be used with the pump.
2.9 Personal Diabetes Assistant
(PDA) P
recautions
The PDA is the main interface for the
entire system. Please note the following
precautions:
Do not let others handle your
PDA, except your qualified healthcare
professional.
Please keep the battery charged .
Do not drop or immerse in water as
this could cause a malfunction.
2.10 Blood Glucose Test
Function Precautions
The blood glucose testing function is
for in vitro diagnostic use only.
The PDA is only compatible with
MicroTech Exactive EQ blood glucose test
strips.
The blood glucose test function is not
intended for use in the diagnosis or
screening of diabetes mellitus, nor for the
use on newborns.
Test results may be inaccurate if
hematocrit levels fall outside the test
strip hematocrit range of 30-55%.
High levels of vitamin C or other
reducing agents can lead to inaccurate
results.
healthcare professional and/or local
distributor, Homemed (Pty) Ltd for support.
may result in unsafe practices. Microtech
Medical and Homemed (Pty) Ltd is not
liable for any related legal obligations
arising from misuse.
Use only fresh whole blood samples
to test your blood glucose.