Equil
TM
Insulin Patch Pump System
105
16.1 Patch Pump Alarms
Alarm Description Priority Level Alarm Signal Solution/Action
No Insulin
Remaining
Blockage
Detected
(Occlusion)
Unexpected
Delivery Stoppage
Pump Battery
Exhausted
Cannot Give Full
Bolus/Dose Low
Pump Battery
Level is Low
Medium
Medium
Medium
Medium
Low
Low
Vibration
LED,
Vibration
LED,
Vibration
LED,
Vibration
LED,
Vibration
LED,
Vibration
There is no more insulin and delivery has
stopped. The pump will automatically
rewind. Replace the reservoir and check
your blood glucose level.
Insulin delivery has stopped. Replace
the reservoir/infusion set and check your
blood glucose level.
Insulin delivery has stopped. Check your
BG level. Replace reservoir/infusion set. If
problem persists, contact the supplier for
guidance.
Insulin delivery has stopped.
The pump will automatically rewind.
Please replace the depleted battery with a
fully charged one.
There is not enough insulin left in the
reservoir to execute the requested bolus.
Prepare a new reservoir to be used after
the old one is
empty.
The pump battery level is below 5%.
Please prepare a fully charged battery for
replacement
If the “Auto-Off” function is enabled,
infusion may automatically stop when
preprogrammed this way. For details,
please refer to 14.2.
Note
"Unexpected delivery stoppage" When
this notice appears, the pump infusion
has been stopped and the pump will
automatically rewind. Please check the
battery connection and the pump status,
Measure your blood glucose level.
Note