117
18.4 Bolus Calculator
18.5 Bolus Delivery
18.6 Infusion Precision
At 1U/h delivery rate, the error was
measured at 0.4% as shown in Figure 122.
Figure 122
Bolus Steps Volume
per Step
Time Interval
Between Steps
Infusion Speed
per Minute
0.05 U 0.5 μL 1s 3 U
The above test results were achieved
using pump serial number 0A001 and
reservoir batch number G13B25001
Note
Attribute Specification
On/Off, default is off
1-150 g carb/U in 1 g carb/U increments, no default
0.1-16.7 mmol/L/U in 0.1 steps, no default value
On/Off, Default is On
2-6 hrs in 0.5hr increments,no default value
Bolus Calculator
Carb Ratio
Insulin Sensitivity Factor
Negative Correction
Active Insulin Time