Equil
TM
Insulin Patch Pump System
111
are inaccurate or inconsistent with how
you feel.
•
When you suspect that your meter has
been damaged.
•
After cleaning your PDA.
Refer to Section 10.8 for instructions on
how to perform a quality control test.
Storage and Handling
Please review the following storage and
handling instructions:
•
Store the control solution in the
temperature range 5-30°C (41-86°F).
•
Do not refrigerate or freeze the control
solution.
•
If the control solution is cold, do not use
until it has warmed to room temperature
.
•
Do not use expired control solution.
•
Control solution may be used for 6
months after the bottle is opened for the
first time. Please take note of the date that
the bottle was first opened and discard
after 6 months. Do not use beyond the
expiration date.
Control Solution Precautions:
•
For in vitro diagnostic use. The control
solution is for testing only outside of the
body. Do not swallow or inject.
•
Control solution should be shaken
before use.
•
Quality control tests should be carried
The control solution label contains
the expiration date in year-month
format. For example, 2019-01 indicates
that the control solution is valid until
January 2019.
Note