Equil
TM
Insulin Patch Pump System
7
professional if you experience excessively
high or low blood glucose levels, or if you
have any questions about your care.
2.11.2 Static Electricity
The insulin pump system is resistant
to normal levels of static electricity (ESD)
but high levels of shock may cause a
software reset which can interrupt insulin
delivery.
ESD are more likely in situations
where the relative humidity is very low,
such as inside a heated building during
the winter in areas where it is cold
outside. If you suspect that your pump
has failed, follow the instructions listed in
Section 16: Alarms and Troubleshooting.