123
IEC60601-1-2: Table 206
Recommended separation distances between portable and mobile RF
communications equipment and the device
Maximum rated
output power of
transmitter
Separation distance according to frequency of transmitter (m)
150kHz~80MHz
d=1.2
P
80MHz~800MHz
d=1.2
P
800MHz~2.5GHz
d=2.3
P
These devices are intended for use in an environment in which radiated RF disturbances are
controlled. The user of the device can help prevent electromagnetic interference by maintaining a
minimum
distance between portable and mobile RF communications equipment (transmitters) and
the device as recommended below, according to the maximum output power of the communications
equipment
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter ma
nufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23