Place the calibration posiblock in the blocking head.
Press to confirm.
> The blocking arm does a series of movements above the Hartmann plate.
The test is successful: press to go back to the work screen.
The test has failed: run the calibration cycles for the Hartmann plate and centerer .
Calibrating the Hartmann plate
Select the calibration .
Press to start the cycle.
Wait : the points of the plate light up.
> : The calibration is finished.
Calibrating the height under the lens
Select the calibration .
Press to start the cycle.
Wait : the blocking arm performs a series of vertical movements to recalculate the height under the lens.
> : The calibration is finished.
Calibrate the blocker
Select the calibration .
Press to start the cycle.
> An information message is displayed.
Place the calibration posiblock on the blocking head.
Press to confirm.
Wait : the blocking arm performs a series of movements.
> : The calibration is finished.
Calibrate the centering device
Select the calibration .
Press to start the cycle.
> An information message is displayed.
Place the calibration posiblock on the blocking head.
Press to confirm.
Wait : the blocking arm performs a series of movements above the Hartmann plate.
> : The calibration is finished.
Once the calibration is finished, you can re-start the and test cycles to check it.
If the calibration fails, restart the cycle. If it fails again, contact the technical department.
c. Calibrating the touch screen
Calibrate the touch screen if your operations have become difficult due to lack of precision.
Touch screen malfunction
You can directly access the calibration screen without using the settings menu. To do this, press on the centering screen for
5 seconds.
Calibrating the touch screen