USER MANUAL > CENTERING A LENS
Mr Blue 2.0 > v4.0 -09.16 57
1.
2.
3.
Display of the optical center of the lens (white cross) and boxing center of the shape (blue cross)
Display of the sphere power
Display of the cylinder power
Display of the cylinder angle
New manufacturer marking
If the marking is not recognized and the automatic centering fails, we advise you to record the marking of the lens (in
“Interview” mode) in the embedded base so that it can be recognized automatically on next centering. Centering in
“Interview” mode will allow you to compel the use of embedded customer base.
For more information, refer to the section .Centering a lens > Markings database (p.77)
Lens fit check
A visual warning enables you to check the lens fit: any part of the lens outside the shape is displayed in red.
You can modify the PD and pupillary height to get the shape to fit into the lens, but if the induced prism is too great, we
advice you to order a new lens in the proper diameter.
To carry out lens blocking, refer to the section .Centering a lens > Blocking a lens (p.80)
c. Centering a progressive lens in manual mode
Manual centering enables you to center the lenses which:
cannot be centered in automatic mode via re-marked micro-engravings
cannot be centered automatically using the three focimeter dots
Description of the centering target
Centering cross (white cross) and boxing center of the shape (blue cross)
Near vision mark
Centering marks
The centered lens is displayed in the work area:
>
The lens is centered and ready to be blocked.>