U2836 Operation Manual Chapter 5 Execute LCR operation and some examples
42
7) If the test result is obviously incorrect, please check the following items.
a) Check the tested inductance is in good connection with the test fixture or not.
b) Check the test fixture is in good connection with the test terminals of the
instrument or not.
c) Redo the open/short correction.
ÅNOTE: When the sweep open/short correction is used, the point-frequency
correction function should be set as OFF. Refer to Correction operation in this
chapter.