U2836 Operation Manual Chapter 5 Execute LCR operation and some examples
44
6.1.7 Delay time
Delay time: time from trigger to start. 0 to 60s are programmable with a
resolution of 1ms.
6.1.8 Connection modes of test terminals
U2836 adopts 4-terminal test method.
HD(Hcur): current sample high terminal
LD(Lcur): Current sample low terminal
HS(Hpot):voltage sample high terminal
LS(Lpot): Voltage sample low terminal
6.1.9 Test speed (Frequency>=10kHz)
Fast:about 40 times/s(27ms/time)
Medium:about 10 times/s(90ms/time)
Slow:about 3 times/s(300ms/time)
The fast and middle speed will be slow down when frequency <10kHz.
6.1.10 Average
Be programmable from 1 to 255.
6.1.11 Display digit
6 digits, max. displayed digit: 999999
6.2 Test signal
6.2.1 Test signal frequency
Test signal is sine wave, accuracy:0.01%
Frequency range:50Hz~200kHz
6.2.2 Signal mode
Normal: When testing, on measurement display page, voltage across test
terminals may be smaller than preset voltage.
Constant level: The auto adjustment of internal level makes the voltage of DUT
accordant with preset voltage.