VISUAL AND ACOUSTIC ALARM
• Air condition the environment.
• Clean the condenser (see par. 6.3)
• Check that installation has been properly performed (see par. 4.2)
• Switch-off the equipment from keypad (see par. 5.2.3.1) and then,
switch it on again.
• Check that door is properly closed.
• Switch-off the equipment from keypad (see par. 5.2.3.1) and then,
switch it on again.
• Check ambient conditions (environment should be neither too hot
nor too cold).
• Check that installation has been properly performed (see par. 4.2)
• If the problem persists, contact Technical Support.
• Check that products have been correctly stored in the chamber (see
par. 4.6.3)
• Check ambient conditions (environment should be neither too hot
nor too cold).
• Clean the condenser (see par. 6.3)
• Check that door is properly closed.
• Air condition the environment.
• Contact Technical Support.
• Check the electrical system or accidental disconnection of the plug.
• Replace button battery on electronic board.
NO BATTERY (only if DMLP
Digital Monitor is installed)
BATTERY FAILURE (only if DMLP
Digital Monitor is installed)
• Contact Technical Support. Refrigerator functioning is guaranteed
anyway.