A Menu Structure FLUXUS F70x
UMFLUXUS_F7V4-6-2EN, 2017-10-01 173
enabling of the NoiseTrek mode x
Selection if the toggling between the TransitTime and the NoiseTrek mode has
to be carried out manually or automatically.
This display will only be indicated if the NoiseTrek mode is enabled.
x
Input of the time after which the transmitter has to toggle to the NoiseTrek mode
if there are no valid measured values in the TransitTime mode.
range: 0...9999 s
0: no toggling to the NoiseTrek mode
This display will only be indicated if the automatic toggling between the Transit-
Time and the NoiseTrek mode is activated.
x
Input of the time after which the transmitter has to toggle to the TransitTime
mode if there are no valid measured values in the NoiseTrek mode.
range: 0...9999 s
0: no toggling to the TransitTime mode
This display will only be indicated if the automatic toggling between the Transit-
Time and the NoiseTrek mode is activated.
x
Input of the time after which the transmitter has to toggle to the TransitTime
mode.
range: 0...9999 s
0: no toggling to the TranitTime mode
This display will only be indicated if the automatic toggling between the Transit-
Time and the NoiseTrek mode is activated.
x
Input of the time after which the transmitter has to toggle to the NoiseTrek mode
if there are no valid measured values in the TransitTime mode.
range: 0...9999 s
This display will only be indicated if the automatic toggling between the Transit-
Time and the NoiseTrek mode is activated.
x
activation of the display for the difference between the measured and the ex-
pected sound speed of a selected reference medium during the measurement
x
selection whether the flow velocity is displayed and transmitted with or without
profile correction
x
input of an upper limit of the flow velocity
range: 0.1...25.5 m/s
0 m/s: no detection for outliers
All measured values that are greater than the limit will be marked as outliers.
x
selection of the input of a lower limit for the flow velocity:
absolut: independent of the flow direction
sign: dependent on the flow direction
x
activation of the input of a lower limit of the flow velocity:
factory: the default limit of 2.5 cm/s will be used
user: input of a limit
x
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s (0.127 m/s)
default: 2.5 cm/s (0.025 m/s)
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
INIT-
resistant
Enable NoiseTrek
off >ON<
Auto NoiseTrek ?
no >YES<
TT-Failed |After
NoiseTrek | 40s
NT-Failed |After
TransTime | 60s
NT-Ok,but | Each
check TT | 300s
Keep TT | For
checking | 5s
Compare c-fluid
no >YES<
Flow Velocity
normal >UNCORR.<
Velocity limit
0.0 m/s
Cut-off Flow
absolut >SIGN<
Cut-off Flow
factory >USER<
+Cut-off Flow
2.5 cm/s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F705**-NN Series and is the answer not in the manual?

Flexim FLUXUS F705**-NN Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F705**-NN Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals