EasyManua.ls Logo

Flexim FLUXUS G601 - Alarm Output Configuration Details

Flexim FLUXUS G601
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
174 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x A Menu Structure
input of the error value delay, i.e. of the time in-
terval after which the value entered for the error
output will be transmitted to the output if no val-
id measured values are available
This display will only be indicated if Special
Funct.\SYSTEM settings\Dialogs/
Menus\Error-val. delay = EDIT has been
selected.
Pulse Output
Activation of a Pulse Output
This display will only be indicated if a pulse out-
put has been installed in Special
Funct.\SYSTEM settings\Dialogs/
Menus\Proc. outputs.
input of the pulse value (value of the totalizer at
which a pulse will be emitted)
This display will only be indicated if Pulse
Output is activated.
input of the pulse width
range: 1...1000 ms
This display will only be indicated if Pulse
Output is activated.
Alarm Output
activation of an alarm output
This display will only be indicated if an alarm
output has been installed in Special
Funct.\SYSTEM settings\Proc. out-
puts.
Selection of the switching condition (func), the
holding behavior (typ) and the switching func-
tion (mode) of the alarm output.
This display will only be indicated if Alarm
Output is activated.
cold start
resistant
Error-val. delay
10 s
Pulse Output
B1: no >YES<
Pulse Value
0.01 m3
Pulse Width
100 ms
Alarm Output
no >YES<
R1=FUNC<typ mode
Function: MAX

Table of Contents

Related product manuals