198 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x B Technical Data
Shear Wave Transducers (extended temperature range)
technical type GDM2EZ7 GDP2EZ7
order code GSM-ENNNL GSP-ENNNL
transducer
frequency
MHz 1 2
medium pressure
1
min. extended metal pipe: 20 metal pipe: 20
min. metal pipe: 30
plastic pipe: 1
metal pipe: 30
plastic pipe: 1
inner pipe diameter d
2
min. extended 30 15
min. recom-
mended
40 20
max. recom-
mended
80 40
max. extended 120 60
pipe wall thickness
min. 2.5 1.5
max. - -
material
housing PI with stainless steel
cap 304 (1.4301)
PI with stainless steel
cap 304 (1.4301)
contact surface PI PI
degree of protec-
tion according to
EN 60529
IP 65 IP 65
transducer cable
type 6111 6111
length 4 4
dimensions
length l 62.5 62.5
width b 32 32
height h 40.5 40.5
dimensional
drawing
operating temperature
min. °C -30 -30
max. °C +200 +200
temperature
compensation
xx
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
2
shear wave transducers:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a
flow velocity of
bar
bar
mm
mm
mm
mm
mm
mm
m
mm
mm
mm
l
h
b
l
h
b
15 m/s

Table of Contents