18 Settings
FLUXUS G801 18.1 Dialogs and menus
153
UMFLUXUS_G801V1-0EN, 2020-06-26
18.1.2 Fluid pressure
The dependence of the parameters of a fluid on the pressure can be taken into account.
Select on to request the fluid pressure in the program branch Parameter. Select off if 1 bar is to be used for all calculations.
This display will only be indicated if Special Funct.\SYSTEM settings\Measuring\Gas-Measuring is deactivated.
If Gas-Measuring is activated, the fluid pressure will always be requested in the program branch Parameter.
Press ENTER.
18.1.3 Measuring point number
Select (1234) if the measuring point is to be identified only by numbers, point and hyphen.
Select (↑↓← ) if the measuring point is to be designated with ASCII characters.
Press ENTER.
18.1.4 Transducer distance
recommended setting: user
Select user if the measuring point is always the same.
auto can be selected if the measuring point often changes.
In the program branch Measuring, the recommended transducer distance is displayed in parentheses, followed by the
entered distance if the 2 values differ from each other.
During the transducer positioning, in the program branch Measuring:
only the entered transducer distance will be displayed if Transd. Distance = user is selected and the
recommended and the entered transducer distances are identical
only the recommended transducer distance will be displayed if Transd. Distance = auto is selected
18.1.5 Error delay
The error delay is the time interval after which the error value is transmitted to the output in case no valid measured values
are available.
Select damping if the damping factor is to be used as the error delay. Select edit to activate the menu item
Error-val. delay in the program branch Output Options. From now on, the error delay can be entered
in the program branch Output Options.
Press ENTER.
Special Funct.\...\Dialogs/Menus\Fluid pressure
Notice!
For documentation purposes, it is useful to enter the fluid pressure, even if the transmitter contains no pressure-
dependent characteristic curves.
Special Funct.\...\Dialogs/Menus\Meas.Point No.
Special Funct.\...\Dialogs/Menus\Transd. Distance
Special Funct.\...\Dialogs/Menus\Error-val. delay
Transd. Distance
(50.8) 50.0 mm