5. Operation 
Note: Before operating the device, please read, understand, and follow all instructions, 
dangers, warnings, cautions, and notes. 
 
Note: When the meter is not in use, the function switch should be set to the OFF 
position. 
 
Note: When connecting the probe leads to the device under test, connect the negative 
lead before connecting the positive lead. When removing the probe leads, remove the 
positive lead before removing the negative lead. 
 
5.1 Powering the Meter 
1. Set the function switch to any position to switch the meter ON. 
2. If the battery indicator   shows that the battery voltage is low or if the meter does not 
power on, replace the batteries. See section 6.2 Battery replacement.  
 
5.1.1 Auto Power Off  
The meter enters sleep mode after 10 minutes of inactivity. The meter beeps three 
times before powering off. Press any button or turn the function switch to prevent 
the meter from powering off. The auto power off time-out is then reset. Deactivate 
APO (or set the APO timer) using the Programming Menu (refer to Section 5.10.2 
Programming Menu). The APO icon   is displayed when APO is active. 
 
5.2 Auto/Manual Range Mode 
In Auto range mode, the meter automatically selects the most appropriate measurement 
scale. In Manual range mode, the desired range (scale) can be set manually. 
Auto range mode is the default mode of operation. When a new function is selected with 
the function switch, the starting mode is Auto range and the   indicator is displayed. 
To enter Manual range mode, press the RANGE button. To change the range, press the 
RANGE button repeatedly until the desired range is displayed. 
To return to Auto range mode from Manual mode, press and hold the RANGE button 
until the   indicator is displayed. 
 
5.3 HOLD Mode 
  In Hold mode, the display freezes the last reading and continues to display this 
value. 
  Use the HOLD button to toggle between Normal and Hold modes. In Hold mode, 
the   indicator is displayed.