7. Specifications
7.1 General specifications
Display counts: 0~6000
Measuring rate: 3 times per second
Over-range indication: OL or –OL.
Auto power off: Programmable: OFF, 1, 2, 5, or 10 minutes
Low battery indicator: and ‘LO BATT’ is displayed.
Power requirement: 4 × 1.5 V AAA alkaline batteries.
Battery life: Clamp mode Approx. 10 hours; IGMTM Mode 3 hours (with
alkaline batteries and work light off)
Calibration: 1-year calibration cycle.
Operating temperature:
14 to 50°F (-10 to 10°C) (non-condensing)
50 to 86°F (10 to 30°C) (≤ 80% RH)
86 to 104°F (30 to 40°C) (≤ 75% RH)
104 to 122°F (40 to 50°C) (≤ 45%RH)
Storage temperature:
4 to 140°F (-20 to 60°C)
Relative Humidity: 0~80% RH (batteries not installed)
Temperature coefficient: 0.2 × (specified accuracy)/°C, <64.4°F (18°C), >82.4°F (28°C)
Operating altitude: 6562ʹ (2000m)
Jaw opening: 1.38in. (35mm)
Pollution degree: 2
Dimensions: (D × W × L): 1.7 × 3.5 × 9.5 in. (43 × 89 × 241 mm)
Weight: 15.0 oz. (426g) including batteries
Drop test: 6.6 ft. (2 meters)
Agency Approvals: UL, CE, RCM, and KC (MSIP-REI-FS5-CM174)
Over-voltage category: EN 61010-1 CAT IV-600 V, CAT III-1000 V, EN 61010-2-032
Installation sources, utility transformers, all outside conductors, counters,
protective devices on primary sides and electricity meters