7.3 Electrical specifications 
Accuracy is ± (% reading + number of digits (dgt)) at 73.4 °F ±9 °F (23°C ±5 °C), <80% RH. 
 
Table 7.1 Voltage (TRMS) 
Range, resolution, and accuracy specifications for Low 
Impedance (Lo Z) measurements are the same as DCV, 
ACV, and VFD specifications shown above 
 
Notes: 
LCD Displays ‘0’ counts when the AC reading is <10 counts 
Overload protection: 1000V (rms) 
Input impedance: 10 MΩ //, <100 pF 
Lo Z input impedance: 2.5kΩ 
AC conversion type: AC coupled, true RMS responding, calibrated to the RMS value of a sine wave 
input. Accuracies are given for sine waves at full scale and non-sine waves below half scale. 
If the meter measures a 4000-count signal and the Crest Factor of the signal is more then 3.0, the 
reading may not meet specified tolerances. For non-sine waves (50/60 Hz), add the following Crest 
Factor corrections: 
For non-sine waves (50/60 Hz), add the following crest factor corrections: 
For a crest factor of 1.0–2.0, add 3.0% to the accuracy. 
For a crest factor of 2.0–2.5, add 5.0% to the accuracy. 
For a crest factor of 2.5–3.0, add 7.0% to the accuracy.